6. Cleaning and Maintenance
Cleaning of the following IP modules signed by
an arrow is of particular importance:
Cassette guiding mechanisms
• Loadingstation(Fig. 54)
The ejection unit of the magazines, magazine
holders and chute.
Always ensure that the sensor at the end of
the chute is clean.
• Transportstation(Fig. 56)
Remove dust and debris from the Cassette-
Clamp.
Fig. 54
Fig. 55
6.1 Cleaning the instrument
Prior to cleaning the instrument, always switch off power supply and unplug the power cord!
When handling cleaning detergents, follow the instructions of the manufacturer and make sure
all laboratory regulations in force in your country are complied with.
To clean the exterior surfaces, use a mild and ph-neutral commercial household cleaner. You
may NOT use: Alcohol, cleaning materials containing alcohol (glass cleaner!), abrasives, or
solvents containing acetone or xylene!
No liquid may come into contact with the electrical connections or spill into the interior of the
instrument!
The Leica IP C needs to be thoroughly vacuum cleaned once a week.