Dil ID: 14
Dil Adı: sloavakia
Dil Kodu: sk7tayfun Produkty značky Leica-biosystems
Dünyanın En Büyük
Elektronik Bilgi Kütüphanesi



45
Leica IP S
5. Operation
5.2 Display indications
Display indication Magazine empty (in combination with "Mag. Empty" LED)
1 - Magazine no. 1 empty
2 - Magazine no. 2 empty
3 - Magazine no. 3 empty
If "MANUAL FEED" has been selected in the printer driver settings, "0" will
appear in the display after the print job has been sent. The printer will wait
for an individual specimen slide to be placed in the feeding chute for printing.
Display
Status messages
00 Ink print head cleaning in progress.
11 Temperature in the flash power supply is too high.
Printer is too hot—there will be a short cool-down period.
The print job will be resumed automatically after a short period of time.
To prevent frequent job interruptions due to heat build-up, make sure
the ventilation grids of the printer are unobstructed and keep the
printer away from other heat sources.
Consider operating the printer in an air-conditioned room.
If the temperature does not drop to a value within the allowed range
within 10 minutes, "55" is displayed. Switch the instrument off and let
it cool; check ambient temperature.
12 Drying module temperature too low/too high.
If the temperature does not drop to a value within the allowed range
within 6 minutes, "43" is displayed.
13 Flash bulb has reached its maximum life.
The flash lamp has reached the end of its specified service life and
must be replaced.
If this message is ignored, the resistance of subsequent printouts
can be affected.