Dil ID: 10
Dil Adı: spain
Dil Kodu: es7tayfun Productos de la Marca Leica-biosystems
Dünyanın En Büyük
Elektronik Bilgi Kütüphanesi



58
Instructions for Use V 1.9 RevD – 05/2013
• Thenpulloutthelocationplate(70, Fig. 64).
• Cleanthe printheadwith a foamswab,as
described in Fig. 59 (see also Fig. 65).
70
• After cleaning the print head,
insert a new transport plate
(73).
Push the small lever (69,
Fig. 58) back down into its
original position.
• Reclosethelidoftheprinter
and press any key; the print
head moves back into the rest
(park) position.
Screwtheredcap (85, in
Fig. 57) onto the cleaning
cartridge and close the plate
on the left side again.
Instructions for storage (continued)
Fig. 64
Fig. 65
89
73
6. Cleaning and Maintenance
Always use the cleaning swab for the
print head once only. Never rotate the
swab—this can damage the nozzle
plate of the print head.
Important!
Turn the printer off and disconnect it from the power supply at this point to prevent damage to
the print head.
Never use a cleaning cartridge together with a location plate (70)!
Do not reinstall a used transport plate (73), as it will no longer completely seal the print head.