
29
KDS EZFlow 2040
Alarm Messages (Continued)
5. EMPTY ALARM
(Continued)
IMPORTANT: The EMPTY ALARM can only be triggered when the drop sensor is plugged
in; The EMPTY ALARM will be triggered only if there are no drops going through the drip
chamber.
WARNING: There are three situations that may trigger the EMPTY ALARM if drop sensor is
used:
a. Medication in the reservoir is finished;
b. Drip sensor is not properly plugged into the pump and/or clamped on the drip-
chamber;
c. Drip sensor is removed from the drip-chamber; in the above three situations,
EMPTY ALARM will be triggered with an audible and visual flash light.
Pump will stop regular infusion and switch to a KVO (Keep Vain Open) infusion
mode. Press to silence the alarm buzzer.
IMPORTANT: Infusion can be resumed ONLY after the EMPTY condition is cleared.
EXIT
CLEAR
6. COMPLETE ALARM
EXIT
CLEAR
Screen 32
IMPORTANT: To restart or clear the total infused volume, press ON/OFF switch on the back
of the pump. After the pump is turned OFF, user must wait at least 3 seconds to turn on the
pump again.
During a regular infusion, when the VOLUME
DELIVERED matches the VOLUME LIMIT (vol-
ume to be infused), COMPETED alarm will be
triggered with audible alarm and visual flash
light. (screen 32) Pump will stop regular infu-
sion and switch to a KVO (Keep Vain Open)
infusion mode (KVO rate is 0.1 ml/h at micro
infusion; 1ml/h or 2 ml/h at macro infusion).
User can press to silence the alarm
buzzer.