Dil ID: 1
Dil Adı: english
Dil Kodu: en4 ZMODO Brand's Products
Dünyanın En Büyük
Elektronik Bilgi Kütüphanesi



3232323232
30
KDS EZFlow 2040
Alarm Messages (Continued)
7. LOW BATTERY ALARM
Screen 33
Screen 34
If battery power is low while pump is operat-
ed with internal battery power supply,
BATTERY LOW ALARM will be triggered
with audible alarm and visual flash light.
(screen 33)
WARNING: If BATTERY LOW ALARM Is triggered, user must connect pump into a
electrical outlet immediately to work under plug-in power and charge the internal
batteries. Failure to do that may cause the pump working abnormally and result
serious injury to the patient!
When the BATTERY LOW ALARM is trig-
gered for the first time, pump will continue
operating for a while depending on the cur-
rent infusion rate. User MUST connect the
pump to the electrical outlet immediately. If
pump is not connected into an electrical outlet
within 30 minutes, PUMP STOPS IN 3
MINS alarm will be displayed on the screen
(screen 34) with audible alarm and visual
flash light. Pump will stop in 3 minutes after
the 3-minutes alarm is triggered.