Multitronic Novo plus controls
12
G 7883 and G 7883 CD
Washer disinfectors
Controls
● Multitronic Novo plus,
some process parameters can be
altered
Programmes
● 8 standard-cleaning programmes
Programme control
● Single knob programme selector
switch
Display
● Programme sequence display
● Temperature and programme dura-
tion
● Programme end, optical and acous-
tic signal
● Service and fault check lights
Control and safety features
● 2 NTC probes in the sump for tem-
perature monitoring and redundant
temperature control
● Electric door lock
● Programme safety cut out
● Programme stopped in event of
power cut
Interface
● Serial interface RS 232 for PC or
printer for process documentation
(This feature is standard or optional
depending on country)
● Optical interface for service techni-
cian
Note
See page 35 for a list of programmes
and their contents
Multitronic Novo plus
Serial interface for PC or printer connec-
tion
Process data reporting using a laptop
computer
Access point for measuring tempera-
ture holding times