It is the responsibility of the operator
to routinely check that the required
cleaning standards are being met.
Therefore, results need to be regularly
tested and documented, both thermo
electrically and through inspection. For
thermo chemical processes, additional
testing is required using chemical or
biological indicators.
~
The performance of the cycles was
tested using Neodisher products. The
use of other products is not
discouraged but may not yield identical
results. Please be aware that changes
in formulation, storage conditions, etc.
which may not be publicized by the
chemical manufacturer can have a
negative effect on cleaning results. and
are not the responsibility of the machine
manufacturer.
~
The process must be set so that no
foam escapes the wash compartment.
Escaping foam jeopardizes the safe
operation of the lab washer.
~
The process must be checked
regularly in order to detect any
foaming.
~
In order to prevent damage to the
lab washer and any accessories
through processing chemicals or dirt as
well as any interaction between them,
see "Effects of Processing Chemicals".
~
When a chemical additive is
recommended, the manufacturer of the
machine takes no responsibility for the
effect of the chemical on the items
being cleaned.
~
The mobile units, baskets and
special inserts should only be used for
their specific applications. Instruments
with hollow parts must be flushed
through with wash water.
~
Empty any containers or utensils
before arranging them in the machine.
~
Do not allow any remains of acids,
solvents or corroding ferrous material,
and in particular hydrochloric acid or
chloride solutions to get into the wash
cabinet. Similarly avoid any materials
with a corrosive effect. The presence in
compounds of any solvents should be
minimal, (especially those in hazard
class A1).
~
To avoid any corrosion damage
ensure that solutions or steam
containing hydrochloric acid do not
come in contact with the stainless steel
casing of the machine.
~
Please follow the installation advice
in these instructions and the separate
Installation Instructions.
IMPORTANT SAFETY INSTRUCTIONS
7