Dil ID: 10
Dil Adı: spain
Dil Kodu: es7tayfun Productos de la Marca Leica-biosystems
Dünyanın En Büyük
Elektronik Bilgi Kütüphanesi



63
Leica IP C
7. Troubleshooting
Error messages (continued)
Display
code
Source of error Troubleshooting solution
Cover (4 in Fig. 2) not closed properly.
Safety switch triggered.
Ensure that the cover is not blocked, e.g. by the
manual unload station. Close cover completely.
Flash standby state not reached within the pre-
scribed time. Charging electronics defective.
Call Leica Technical Service.
No flash or flash duration too short.
Flash bulb dirty or defective.
Check whether flash bulb works – to do so,
observe the stray light on the cover.
Never open the cover to check whether a
flash is triggered!
No flash --> install a new flash bulb. See
Chapter 4.6.
The temperature in the drying module is not
within the permissible temperature range.
Carry out a RESET. See Chapter 7.1.
Unload station mechanically blocked. Remove the cause of the blockage.
Heater fan not running or running too slowly. Call Leica Technical Service.
No cassette is in the drying module for flashing.
Cassette left the cassette carrier, but did not
reach the drying module.
Remove cassette.
Cassette not successfully ejected from drying
module.
Remove cassette.
Cassette in the drying module during initialization
or prior to a print job being carried out.
Remove the cause of the blockage.
The flap of the drying module either does not
close or does not open. The flap is blocked (e.g.
by a cassette).
Remove the cause of the blockage.
Lifting table end sensor does not switch. Call Leica Technical Service.
Lifting table position sensor does not switch dur-
ing positioning.
Call Leica Technical Service.
Flash power supply: Overtemperature for more
than 10 minutes.
Call Leica Technical Service.