STEP 4. Wiring the booster amplifier
Power supply wiring to the booster amplifier
The booster amplifier, whether mounted integrally or remotely with the
D600 sensor, requires a separate AC power supply. To wire the booster
amplifier power supply:
1. Remove screw and terminal cover before installing wiring. Grease
O-ring and re-install cover before operating.
2. Provide 85-250 VAC power to terminal N/L2 and terminal L/L1.
For integral booster amplifier wiring, see Figure 8.
For remote booster amplifier wiring, see Figure 9.
CAUTION
Improper installation of wiring could cause
measurement error or sensor failure.
Shut off power before installing power-supply wiring.
Follow all instructions to ensure sensor will operate
correctly.
Install drip legs in conduit or cable.
Seal all conduit openings.
Ensure integrity of gaskets, grease all O-rings, and
fully tighten sensor junction-box cover, booster
amplifier cover, core processor cover, and all
transmitter housing covers.
WARNING
Improperly removing the booster amplifier housing
cover in a hazardous area could cause an explosion.
In a hazardous area:
Do not open booster amplifier housing cover while
booster amplifier is energized.
Wait at least 30 minutes after power is shut off before
opening.