Screen 20
Operation Settings and Procedures (Continued)
8. ADJ. I.V SET (ACCURACY CALIBRATION
Continued)
MANUAL ADJUSTMENT (For ml/h or drop/min mode)
1) Press to choose manual
calibration. (screen 20)
2) Press or to change calibration
volume in percentage.
3) Press to confirm the adjustment
and return to the MODE SETTING status.
IMPORTANT : The manual adjustment is in effect at the current flow rate from 0.1 to 1000
ml/hr ONLY. The pump will NOT respond to the manual adjustment if the current flow rate is
set from 1001 to 1200 ml/hr.