Protective Ground
Terminal
CAUTION
Refer to Manual
General Information
This Micro-Infusion Syringe Pump is a conventional device for research and development
labs and universities. This Micro-Infusion Syringe Pump offers a reliable alternative for fluid
control, especially suitable for long-term, micro fluid and precise injections.
CAUTION
Important: User should read this entire manual before operating this micro-infusion syringe
pump.
Keep this manual for future reference!
General Safety Summary
WARNINGS:
1) Do not use a pump, syringe, or accessories that show any sign of damage.
2) Check the “dose calibration” switch located on the bottom of the pump every time
use a new/different size/brand syringe. User must calibrate the pump before use
to insure the accuracy of infusion.
3) Only authorized service personnel should repair this pump. Manufacturer assumes
no responsibility for incidents that occur if the pump is not repaired according to
manufacturer-authorized procedures.
4) Before connecting the syringe, user must prime the syringe and
extension tube to prevent infusing air.
5) To clean the pump, use a damp (not wet) cloth or sponge, Do not immerse pump
or power cable into water or other cleaning solution.
6)
Only use the 3-Pin AC (main) power cord included with this pump. Use of any
other cord presents a hazard of shock or death to subject. The AC mains outlet
must be properly grounded to prevent shock hazard. Main outlet and
ground should be tested on a regular basis to ensure safety. Operate away from
strong magnetic fields.
7) This pump will go through a self diagnostic program when the power switch is
turned on. Normally a “0.1” is shown in the display. This indicates the pump is
operating properly. If an “ERR” is shown in the display. reboot the pump. If the
error condition still exists, do not continue operation and contact manufacturer for
technical service immediately.
KDS EZFlow 2010/2011