57
Leica IP S
6. Cleaning and Maintenance
• Thenpulloutthelocationplate(70), (Fig. 60).
• Cleantheprintheadwithafoamswab,asde-
scribed in Chap. 6.2, "Cleaning the print head"
(see also Fig. 61).
70
• Aftercleaningtheprinthead,
insert a new transport plate
(73).
• Pushthe smalllever (69,
Fig. 59) back down into its
original position.
• Reclosethelidoftheprinter
and press any key; the print
head moves back into the
rest (park) position.
• Screw the red cap (85,
Fig. 58) to the shutter, close
it securely, and take the ink
cartridge out completely.
Instructions for storage (continued)
Fig. 60
Fig. 61
89
73
Always use the cleaning swab for the
print head once only. Never rotate the
swab—this can damage the nozzle
plate of the print head.
Important!
Turn the printer off and disconnect it from the power supply at this point to prevent damage to
the print head.
Never use a cleaning cartridge together with a location plate (70)!
Do not reinstall a used transport plate (73), as it will no longer completely seal the print head.