Section 2 System Overview
2-6-2 Turning on the Power Supply
Tum on the power supply of each unit.
When only reading data, the laser power supply and illumination power supply are not
necessary and need not be tumed on.
To obtain a stable laser output, tum on the laser power supply and allow the system to warm
up for 10 min. or more before using.
Once the illumination power supply (mercury vapor lamp power supply) has been tumed off,
do not turn it on again for at least 10 min. to avoid shortening the service life of the mercury
vapor lamp.
• Tuming on the laser power supply
Argon laser:
(1) Turn on the power supply switch.
(2) Turn the key to the ON position.
Krypton/argon laser:
Please refer to the laser
instruction manual.
Please refer to the laser instruction
manual.
(1) Tum on the power supply switch.
(D
2-11