3. Precautions for Use of the System
3 Precautions for Use of the System
Installation
• The FLUOVIEW system will be assembled and setup by qualified technicians. Avoid moving the
system as this may adversely affect the adjustment of the optical system. In case it becomes
necessary to move the system, please consult your Olympus representative.
Olympus shall not be liable for improper adjustments, damages, or other problems occum'ng as a
result of the system being moved by the customer or other unauthorized personnel.
• Pulling the optical fiber in case of the KrAr laser configuration will change the laser output
Never modify by pulling the optical fiber.
• To prevent overheating, leave at least 30 cm of free space t>etween the exhaust opening of the
laser cooling fan and walls or other objects. Laser tube overheating will result in unstable laser
oscillation or the laser oscillation may stop. Excessively high temperature inside the unit may also
result in malfunction and/or damage.
• Openings for a cooling fan for ventilation are provided on the rear panel of the control unit
computer, and the magneto-optical disk unit. To prevent the risk of overiieating and other damage,
the openings should never be blocked. Leave at least 10 cm of free space between the openings
and walls or other objects.
Avoid installation in places exposed to direct sunlight high humidity, high temperatures, and dust
Operating environment: Ambient room temperature 10~35°C; relative humidity 30~80%.
• Avoid installation in places exposed to vibrations. Consult your Olympus representative if the
location is subjected to vibrations.
Handling
• The FLUOVIEW system is a precision instrument. Handle it with care and avoid subjecting it to
sudden or severe impact
• Never excessively bend, strain, or subject the laser fiber to crushing force as this will result in
reduced performance.
• The piezoelectric stage and the piezoelectric nosepiece (option) employ a piezoelectric element
which is fragile. Never subject to impact and ensure that undue force is never applied to these
units.
Precautions for Use
• When the built-in power supply for transmitted light is tumed on for a long time, accumulated heat
may cause metal to stretch and the focus to drift. To achieve highly precise image data, it is
recommended to turn OFF the main switch on the microscope frame.
3-1