Model D and DT Sensors Instruction Manual 19
Installation
Step 4 Wiring
Hazardous area installations
The following warning applies to hazardous area installations
Model D sensor junction box
Most Model D sensors are shipped with a junction box for wiring. There
are two types of junction boxes: one for the D600 sensor, one for all the
other D sensors.
The D600 junction box is different than the junction box for other D
sensors. D600 sensors are also available with a core processor. For
more information on the D600 sensor, see page 24.
• Model DT sensors can use a junction box. For more information on the
DT sensor junction box, see page 20.
For Model D sensors (except the DT series and D600):
If it if not already installed, install the junction box on the sensor,
following the wiring instructions on the junction box.
If possible, install wiring with the junction-box opening pointed
downward, or with a drip leg in the conduit or cable, to reduce the risk
of condensation or excessive moisture in the junction box. See
illustration on page 20.
Next, follow the guidelines in Connecting and shielding 9-wire cable to
wire the sensor to the transmitter.
WARNING
Failure to comply with requirements for intrinsic
safety in a hazardous area could result in an
explosion.
Make sure the hazardous area specified on the sensor
approvals tag is suitable for the environment in which the
sensor is installed. See illustrations, pages 3-6.
For installation in an area that requires intrinsic safety,
refer to Micro Motion hazardous approval installation
instructions.
For hazardous area installations in Europe, refer to
standard EN 60079-14 if national standards do not apply.