Technical data
G 7823 – G 7828
Professional cleaning machine G 7823 (1 door) G 7825 (1 door) PG 8527 (1 door)
G 7824 (2 door) G 7826 (2 door) PG 8528 (2 door)
Front loader G 7823 G 7825 PG 8527
Through feed operation (for clean/unclean side) G 7824 G 7826 PG 8528
Single or side by side installation
l l l
Fresh water system, max. temperature 93°C l l l
Direct coupling of mobile units for processing hollow instruments l l l
2 circulation pumps [max. throughput l/min] 300/500* 300/500* 400/600*
Controls/programmes
Freely programmable PROFITRONIC
l l l
64 programme places l l l
Electrical safety lock l l l
Peak load cut-out l l l
Serial interface for documentation processing l l l
Sensors for automatic mobile unit recognition l l l
Plumbing
1x cold water connection, flow pressure 2.0–10 bar
(200–1000 kPa) 4° dH/0.7 mmol/l
l – –
1x cold water connection, flow pressure 2.0–10 bar
(200–1000 kPa) 4° dH/0.7 mmol/l –
l l
1x hot water connection, flow pressure 2.0–10 bar
(200–1000 kPa) 4° dH/0.7 mmol/l
l – –
1 x hot water connection, flow pressure 2.0–10 bar
(200–1000 kPa) 4° dH/0.7 mmol/l –
l l
1x demin. water connection, flow pressure 2.0–10 bar
(200–1000 kPa)
l – –
1x demin. water connection, flow pressure 2.0–10 bar
(200–1000 kPa) –
l l
3 water inlet hoses,
1
/2” with
3
/4” threaded union l l l
DN 50 drain valve with on-site odour trap required l l l
2 drain pumps with drain hose DN 22 and
on-site odour trap required Optional Optional Optional
Electrical connection
3 N AC 400–415 V 50 Hz
l l l
Heater rating [kW] 9.0 9.0 –
Heating/cabinet [kW] – – 18
Heater/boiler [kW] – – 15
Circulation pump [kW] 0.4/0.8 0.3/0.7 0.7/1.2
Total connected load without TA [kW] 10.2 10 –
Total connected load with electric TA [kW] 10.2 10 20
Fuse rating [A] 3 x 16 3 x 16 3 x 32
* 300 l for spray arms/500 l for direct coupling to mobile unit