Quick start
VEGAPULS 66 standpipe version • 4 … 20 mA/HART - two-wire
36523-EN-121011
Quick start
The quick start enables a quick setup with many applications. You
can nd further information in the respective chapters of the operating
instructions manual.
1. In case of turbulences or strong product movement in the vessel,
long standpipe antennas should be fastened to the vessel wall.
2. The measurement is only possible inside the tube, the standpipe
antenna must hence reach up to the requested min. level
For further information see chapter "Mounting".
1. Make sure that the power supply corresponds to the specica-
tions on the type label.
2. Connect the instrument according to the following illustration
5
1
2
+
( )
(-)
678
4...20mA
2
3
4
1
Fig. 1: Electronics and connection compartment, single chamber housing
1 Voltage supply, signal output
2 For indicating and adjustment module or interface adapter
3 For external indicating and adjustment unit
4 Ground terminal for connection of the cable screen
For further information see chapter "Connecting to power supply".
The instrument is preset by default with the suitable parameters for
"Application", "Min. adjustment" and "Max. adjustment".
For adaptation of the parameter "Medium" you have to proceed as
follows:
1. Go via the indicating and adjustment module to the menu "Setup".
2. In the menu item "Medium"you select the medium of your applica-
tion, for example "Solvent".
1. In the menu "Additional settings", menu item "Damping" you have
to adjust the requested damping of the output signal.
2. Select the output characteristics in the menu item "Current out-
put".
Mounting
Connect electrically
Set parameters
Further steps