Dil ID: 1
Dil Adı: english
Dil Kodu: en3 AGAiT Brand's Products
Dünyanın En Büyük
Elektronik Bilgi Kütüphanesi



1212121212
2: HPX-2000-HP-DUV Specifications
4 000-10000-081-02-201207b
Front Panel
Component Description
Power LED Displays state of main power to lamp.
Xenon ON/OFF
Press and hold the ON/OFF switch for more than 1 second to turn the lamp
on or off.
Status LED
Status of the Xenon lamp:
Lit Xenon lamp is on
Blinking Lamp is in setup mode or cool down mode. The Lamp
cannot be started when the Status LED is blinking.
Off Xenon lamp is off
Shutter Switch
OPEN - Shutter open
CLOSE - Shutter closed
TTL - operation by external TTL- signal (HIGH=OPEN, LOW=CLOSE)
Mechanical Protection
SMA Connector
Covered to protect users from unintentionally looking directly at the beam of
light. Used for SMA connections only. Connect the fiber cable to the HPX-
2000-HP-DUV BEFORE turning the lamp on to avoid unnecessary exposure
to UV radiation. Avoid direct contact with the fiber end.
Always wear proper eye protection when using the HPX-2000-HP-DUV lamp.
Four Recessed
Screws
Unscrew to remove the unit’s housing. See Bulb Replacement
.
Four Recessed Screws
(front and back sides)
Mechanical Protection SMA
Connector
Power LED
Xenon ON/OFF
Status LED
Shutter Switch