26
Instructions for Use V 1.9 RevD - 05/2013
4. Instrument Setup
4.5 Installing the manual unload station
Fig. 13
27
26
28
29
32
The enlarged detail shows
the correct location of the
collar screw once the unload
station has been locked in
place correctly.
30
31
31
4
28
The unload station supplied consists of:
26 - Unload station
27 - Screening plate
28 - Collar screws (3 pcs.)
29 - Slotted screws with washers (2 pcs.)
Install as follows (Fig. 13):
• Openlid(4).
• Thereare5tappedholes(2xitemno.30 and
3 x item no. 31) in the installation surface
located below the reflector.
• Withascrewdriver,insert3collarscrews(28)
in tapped holes (31) as far as they will go.
• Then,fastencover (27) in tapped holes (30)
using the two slotted screws (and washers)
(29).
• Tofastentheunloadstationtotheinstrument,
place the wider end of the three oblong holes
(32) over the heads of the three collar screws
(28). Press the unload station against the in-
stallation surface, pushing it simultaneously to
the right until it locks in place (see enlarged
detail).
If the unload station does not easily slide past
the shielding cover, slightly lift the front end of
the device.
• Closelid,makingsuretheunloadstationdoes
not obstruct the lid.
When working with the manual un-
load station the slides imprinted must
be removed in regular intervals. Oth-
erwise they will start piling up in the
unload station, printing will stop and
error code "44" will be displayed.