Instructions for Use V 1.9 RevD – 05/2013
3.3 Print specifications
Only standard histology cassettes that are reli-
ably inserted into the magazines can be used in
the Leica IP C.
3. Instrument Components and Specifications
Fig. 5
3.3.1 Requirements for cassettes
A wide variety of standard cassettes may be
used in the IP C; however, the following restric-
tions must be observed:
• Suitableforprintingareallstandardcassettes
without lids, of the following dimensions:
Length without lid x Width = max. 41.3 x
28.9 mm
Length with lid x Width = max. 80.0 x 28.9 mm
(Refer also to Technical Data, p. 11)
• Cassetteswithlidsattachedhavetobeone-
piece units (23); the lids must not just be linked
to the body by a hinge.
• Cassetteswithaflexiblehingecannotbeused
unless the lid is detached (25) or closed.
Cassetteswith a closed lid(26) must be
checked to ensure that all four corners of the
lid are firmly closed and flat.
• Cassetteswithtop-mountedlids(24) cannot
be processed.
Fordetailsonhow to correctlyinsertthe
cassettes into the magazines, please refer to
chapter 4.7.
Fig. 7