36
Instructions for Use V 1.9 RevD – 05/2013
Fig. 45
87
86
4. Instrument Setup
Exchanging the cartridge (continued)
Fig. 42
Fig. 44
Fig. 43
65
66
• Removetheinkcartridge(66) from the delivery
carton and shake it a few times, then remove
the protective foil.
• Theink cartridge mustbereplaced after
60,000 printouts or 3.5 months. The instrument
setup date can be recorded on the white sur-
face (65) on the front (Fig. 44).
• Insertthecartridgeintotheslot.Loosenthe
red screw-on cap one full turn but do not
remove it yet.
• Insertthecartridgecompletelyintotherecep-
tacleapplyingmoderatepressure.Youhave
to use a certain amount of force to pierce the
seals.
• Withdraw the cartridge from its slot by about
1,25 inches (30 mm) and push it back in as
far as it will go; repeat this step two or three
times.
• Nowunscrewthered screw cap (87) com-
pletely from the nozzle of the cartridge and
store it in the recess provided in the cartridge
for this purpose ( Fig. 43, 44).
This step is very important to ensure
safe operation of the printer.
Very important!
Prior to each transport, the cap (87)
must be screwed onto the nozzle to
prevent the ink from spilling.
• Rightnexttotheslotopeningthereisasafety
catch (86, Fig. 45). Moving this catch to the side
secures the ink cartridge against falling out.